This clinical investigation is designed to evaluate whether respiration rate and heart rate from pediatric patients (5-18 years) can be measured accurately using thoracic motion dataThis study is designed to collect data for testing of theā¦
ID
Source
Brief title
Condition
- Other condition
Synonym
Health condition
vital signals monitoring
Research involving
Sponsors and support
Intervention
Outcome measures
Primary outcome
Primary objective:
The primary objective is obtaining motion sensor data for performance
validation of a respiration rate and pulse rate monitoring algorithm for use on
5-18 years old patients.
The data collected using the motion sensor of this study will be compared with
the one obtained using the golden standard for respiratory frequency ( by
using a respiration band) and heart rate (by using ECG) .
Secondary outcome
The secondary objective is obtaining motion sensor data for the development and
improvement of respiration rate and pulse rate detection algorithms specific
for 5-18 years old patients.
Background summary
Hospital scoring systems like the Early Warning Score (EWS) offer a simple
guide for nursing & medical staff to quickly determine the degree of illness of
a patient by combining their vital signs into one single composite score. It
has been shown that using such scoring system on general wards can reduce the
number of unexpected intensive care admissions by 70% thanks to early detection
of patient deterioration. On pediatric wards, the use of Pediatric Early
Warning Scoring (PEWS) has the potential to greatly enhance patient safety.
The patient*s heart rate and respiration rate are two of the vital signs that
are included in the PEWS system. Measuring these parameters from 5-18 years old
patients multiple times a day requires significant nurse effort, patience of
the patient (sit still during counting) and/or involves wired monitoring. A
method that would allow getting an automated reading on both the heart rate and
respiration rate in a wireless and unobtrusive manner would make the life of
both nurses and patients easier and more comfortable, while improving the
patient*s safety by monitoring their vital signs.
Study objective
This clinical investigation is designed to evaluate whether respiration rate
and heart rate from pediatric patients (5-18 years) can be measured accurately
using thoracic motion data
This study is designed to collect data for testing of the respiration rate and
pulse rate detection algorithm, implemented in the Philips IntelliVue CL
Respiration Pod, for use with 5-18 years old patients. The data collected using
the motion sensor of this study will be compared with the one obtained using
the golden standard for respiratory frequency ( by using a respiration band)
and heart rate (by using ECG) .
Study design
A total number of two investigational devices will be used in this study.
These are: the Respiration pod prototype with the datalogger and the slightly
modified transducer band. The Respiration pod prototype contains the motion
sensors, while the datalogger stores the acquired data from the motion sensors,
transducer band and ECG. The transducer band is a slightly modified version of
a commercially available, CE-approved device. The modification concerns the
connection of pressure studs for connecting the band to the datalogger.
Study burden and risks
Wearing the measurement apparatus may cause some discomfort.
All risks are mitigated to acceptable levels.
Further information can be found in the chapter 11 of the Research protocol.
High Tech Campus 34-1
Eindhoven 5656AE
NL
High Tech Campus 34-1
Eindhoven 5656AE
NL
Listed location countries
Age
Inclusion criteria
Patients of MMC- pediatric department
-Age between 5 and 18 years old
-Subjects must be able to provide an informed consent or have legally authorized representative consent to participate
-Subjects of 12 years and older must provide assent to participate
-Subjects are hospitalized patients of the pediatric department of MMC and stay there for at least 24 hours.
Exclusion criteria
Exclusion criteria for subject selection
* Subjects with deformations that prevent proper device positioning
* Subjects with arrhythmia
* Subjects who had rib surgery
* Subjects with skin irritation that prevent 24 hours wearing the device
* Subjects having implants or devices connected that might be affected by the bioimpedance measurement.