Validation of Microlife 3BTO-A and OMRON for blood pressure measurement in women with moderate to sever preeclampsia.
ID
Source
Brief title
Condition
- Maternal complications of pregnancy
Synonym
Research involving
Sponsors and support
Intervention
Outcome measures
Primary outcome
For each measurement pair the absolute value of the difference [test device] -
[mean investigators auscultatory measurement] is calculated.
Each measurement pair is graded as A: Difference within 5 mmHg , B: difference
within 10 mmHg and C: difference within 15 mmHg.
Depending on the number of readings that fall between pre-specified limits the
device is approved yes or no. The International Protocol specifies that 33
patients are necessary for validation.
If the Microlife 3BTO-A and / or the OMRON meet the requirements of the
International Protocol than use of these devices for blood pressure measurement
can be recommended in women with moderate to severe preeclampsia. If
requirements are not met than these devices should not be used to diagnose
preeclampsia or to initiate treatment or adjust treatment in women with
preeclampsia.
Secondary outcome
nvt
Background summary
Gestational hypertension is on of the most common medical disorder in pregnancy
and occurs in 10-12% of all pregnancies. Accurate measurement of blood pressure
is essential in prenatal care for risk assessment of pregnant women and the
diagnosis of pregnancy induced hypertension and preeclampsia.
The auscultatory technique with a mercury sphygmomanometer is considered the
method of choice, using the first and fifth phases of Korotkoff sounds. [1]
However, there is increasing evidence that this procedure may lead to
misclassification due to patient related, investigator related and device
related factors. Automatic readings may overcome these disadvantages. A number
of automatic devices have been validated in pregnant women. Although the
majority of these devices perform well during normal pregnancy, differences
between automatic reading and auscultatory measurement were large and
unpredictable with most devices in women with preeclampsia.[3-5] Only two
automatic devices (Omron MIT and Microlife 3BTO-A ) have demonstrated
sufficient accuracy for use in women with preeclampsia. [6,7] However, most
women in these studies had only moderately elevated blood pressure and concern
regarding the accuracy at higher blood pressure levels remains. Accuracy at
higher levels in pregnancy is of great importance as medical treatment in
pregnancy is generally considered at higher blood pressure levels than in
non-pregnant individuals. We therefore decided to validate both Microlife
3BTO-A and OMRON in women with moderate to sever preeclampsia
Study objective
Validation of Microlife 3BTO-A and OMRON for blood pressure measurement in
women with moderate to sever preeclampsia.
Study design
The validation procedure will be performed according to the international
protocol for validation of blood pressure devices in adults. [2]. Alternate
auscultatory and automatic readings of systolic and diastolic blood pressure
will be performed. The auscultatory measurements will be performed by two
researchers. The mean value of each pair of investigator measurements will be
calculated. In total, 9 auscultatory and 8 automatic blood pressure readings
will be obtained. The first measurement set of each run is discarded from
analysis.
Study burden and risks
The repeated blood pressure measurements will give some discomfort to the right
arm of the patient. There is no risk involved for the baby or for the mother*s
health.
References
1. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN et
al. Recommendations for blood pressure measurement in humans and experimental
animals: part 1: blood pressure measurement in humans: a statement for
professionals from the Subcommittee of Professional and Public Education of the
American Heart Association Council on High Blood Pressure Research. Circulation
2005;111:697-716.
2. O'Brien E, Pickering T, Asmar R, Myers M, Parati G, Staessen J et
al. Working Group on Blood Pressure Monitoring of the European Society of
Hypertension International Protocol for validation of blood pressure measuring
devices in adults. Blood Press Monit. 2002;7:3-17.
3. Shennan AH, Halligan AW. Korotkoff Sounds. Blood Press Monit.
1996;1:495.
4. Koenen SV, Franx A, Oosting H, Bonsel GJ, Bruinse HW, Visser HA.
Within-subject variability of differences between conventional and automated
blood pressure measurements in pregnancy. Eur.J.Obstet.Gynecol.Reprod.Biol.
1998;80:79-84.
5. Reinders A, Cuckson AC, Jones CR, Poet R, O'Sullivan G, Shennan AH.
Validation of the Welch Allyn 'Vital Signs' blood pressure measurement device
in pregnancy and pre-eclampsia. BJOG. 2003;110:134-38.
6. Golara M, Benedict A, Jones C, Randhawa M, Poston L, Shennan AH.
Inflationary oscillometry provides accurate measurement of blood pressure in
pre-eclampsia. BJOG. 2002;109:1143-47.
7. Reinders A, Cuckson AC, Lee JT, Shennan AH. An accurate automated
blood pressure device for use in pregnancy and pre-eclampsia: the Microlife
3BTO-A. BJOG. 2005;112:915-20.
Meibergdreef 9
22660 1100DD Amsterdam
Nederland
Meibergdreef 9
22660 1100DD Amsterdam
Nederland
Listed location countries
Age
Inclusion criteria
hospital admission because of severe preeclampsia = blood pressure > 140/95 mmHg and proteinuria > 0.3 g/24 hours
Exclusion criteria
None
Design
Recruitment
Followed up by the following (possibly more current) registration
No registrations found.
Other (possibly less up-to-date) registrations in this register
No registrations found.
In other registers
Register | ID |
---|---|
CCMO | NL11954.018.06 |