Primary objective: This is a Post Market Clinical Follow-up (PMCF) study designed to establish the continued safety and effectiveness profile of the Boston Scientific Cardiac Cryoablation System after receiving CE mark.Secondary objective: The study…
ID
Source
Brief title
Condition
- Cardiac arrhythmias
Synonym
Research involving
Sponsors and support
Intervention
Outcome measures
Primary outcome
Primary Saferty Endpoint:
Safety event free rate at 12 months post-index procedure.
Primary safety events will consist of a composite of the following
procedure-related and device-related adverse events.
Acute primary safety endpoint events, events occurring up to 7 days postindex
procedure or hospital discharge, whichever is later, include:
• Death
• Myocardial infarction (MI)
• Persistent gastroparesis/injury to vagus nerve
• Transient ischemic attack (TIA)
• Stroke/Cerebrovascular accident (CVA)
• Thromboembolism/Air embolism*
• Cardiac tamponade/perforation
• Pneumothorax
• Serious vascular access complications**
• Pulmonary edema/heart failure
• AV block not attributable to medication effect or vasovagal reaction.
* Thromboembolic or /air embolic events collected in the study refer to any
occlusion of
blood vessel(s) that results in clinical symptoms
** Defined as prolongation of hospitalization, requirement of surgical
intervention or
blood transfusion
Chronic primary safety endpoint events, events occurring through 12
months post-index procedure, include:
• Atrial esophageal fistula
• Pulmonary vein stenosis (>= 70% reduction of diameter)
• Symptomatic pericardial effusion
• Persistent Phrenic nerve injury***
***A non-recovered phrenic nerve injury at 12 months post-index procedure will
count
as a chronic primary endpoint. The study will collect information on phrenic
nerve palsy
observed before the end of the index procedure and, in case it occurred, will
track
information for potential recovery during the study visits.
Primary Effectiveness Endpoint:
Failure free rate at 12 months post-index procedure.
Failure defined as:
• Failure to achieve acute procedural success* in the index procedure;
• More than one repeat procedure during the blanking period (within 90 days
post-index procedure);
• Documented atrial fibrillation, or new onset of atrial flutter or atrial
tachycardia event (>= 30 seconds in duration or from a 10 second 12-lead ECG)
between days 91 and days 365 post-index procedure **;
• Any of the following interventions for atrial fibrillation, or new onset of
atrial flutter or atrial tachycardia between days 91 and days 365 post-index
procedure:
• Repeat procedure;
• Electrical and/or pharmacological cardioversion for AF/AFL/AT;
• Prescribed a higher dose of any AAD*** documented at baseline or a new AAD***
not documented at baseline.
* Acute procedural success is defined as isolation of the all pulmonary veins
or anatomical equivalents achieved with the Cryoablation Catheter at the end of
the index procedure and as demonstrated at minimum by entrance block using the
Cryo Mapping Catheter (Other techniques of assessment are per investigator*s
discretion).
** Subjects will be monitored for recurrences of the arrhythmias by means of
clinical visits, ECG and 24-hour Holter monitoring.
*** AADs for endpoint will consist of all Class I/III and any Class II/IV
medications taken for control of AF/AT/AFL recurrence
Secondary outcome
Documentation and rate of acute procedural success, defined as pulmonary vein
isolation achieved with the Cryoablation System. Electrical isolation of a PV
is demonstrated at minimum by entrance block using the Cryo Mapping Catheter
(Other techniques of assessment are per investigator*sdiscretion).
Information on isolation method used per site standard of care will be
collected with a minimum by entrance block in all PVs or anatomical
equivalents. Additional data, if available, will be collected on Exit Block or
Conduction Block verification using 3D mapping.
Additional endpoints:
Additional endpoints and analyses include, but are not limited to:
• Procedure times: LA dwell time*, total ablation time, number of cryo
applications per vein, time to thaw, total fluoroscopy time and total procedure
time;
• Time-To-Isolation per ablation application, if available;
• Freedom from recurrence of individual types of atrial arrhythmias (AF, AFL,
AT) between days 91 and days 365 post-index procedure;
• Analysis of ablation techniques (i.e. segmental approach to left common
trunk, additional linear ablations, etc.);
• Analysis of different anaesthesia techniques (General anaesthesia with or
without intubation versus sedation);
• Freedom from recurrence in subjects with non-common anatomical configurations
of PV (e.g. left common trunk);
• Freedom from primary effectiveness failure evaluated in subgroups of subjects
(termination of AAD versus continuation of AAD after blanking period);
• For the subgroup of subjects undergoing 3D mapping with the Boston Scientific
mapping Rhythmia system and performing a post-procedural map, map information
will be collected to determine lesion locations;
• For subjects that will undergo repeat ablation during the course of the
clinical follow-up, anatomical location of ablations gaps will be assessed.
* LA dwell time is defined as time from the Cryoablation Catheter introduced in
the left atrium (LA) to the last time of Cryoablation Catheter exiting the LA.
Additional ancillary analyses on specific subgroup of subjects, may be
presented.
These analyses will include but not limited to the following:
• Age (<60 versus >= 60 years);
• Gender (Male versus Female).
Background summary
Atrial fibrillation (AF) is the most commonly encountered sustained cardiac
arrhythmia in
clinical practice. AF causes symptoms that impair quality of life, increases
the risk of stroke fivefold, and increases mortality. There are multiple
therapies in current use for the treatment of AF like ablation.
Cryoablation has gained significant popularity and utilization worldwide. With
the understanding the pulmonary veins may be the "cornerstone" of ablation
strategies, a cryo balloon has been developed to provide a "single shot"
therapy for isolation of the pulmonary veins. By navigating the balloon to the
ostium of the PV and occluding flow, a PV may be isolated with a single
cryoablation of 3 - 4 minutes.
Complications arising from cryoablation are consistent with those of heat-based
therapies. Additionally, as the balloon is placed in the right-sided PVs and
near the phrenic nerve, diaphragm paralysis, (both transient and permanent),
has been reported. To mitigate this risk, pacing maneuvers and continuous
analysis of diaphragmatic movement has been used.
Study objective
Primary objective: This is a Post Market Clinical Follow-up (PMCF) study
designed to establish the continued safety and effectiveness profile of the
Boston Scientific Cardiac Cryoablation System after receiving CE mark.
Secondary objective: The study will provide information on real-world usage of
the Boston Scientific Cardiac Cryoablation System when used to
perform pulmonary vein isolation (PVI) for the ablation treatment of atrial
fibrillation (AF), according to the current and future guidelines and system
indications for use. This may include but not limited to: repeated ablations to
treat AF, concomitant or delayed adjunctive ablation strategies with
other products and use of different diagnostic products to validate the results
such as 3D mapping systems. Collected information and analyses will include,
but not limited to, the following:
• Arrhythmia characterisation at the time of enrollment: time since first AF
diagnosis, number and duration of reported episodes, prior electrical
cardioversions (if any);
• History of prior ablations and repeated ablations during follow-up period;
• Use of anti-arrhythmic drugs (AAD);
• Methods used for verification of procedure success (pulmonary vein
isolation): entrance/exit block, provocative agents, use of three-dimensional
(3D) mapping.
Study design
Prospective, non-randomized, multicenter (international), single arm study. All
subjects signing the consent, undergoing the index procedure and reated with
the study devices will be followed up for one year. Men and women with the age
of 18 years or above and plannend for a de novo ablation for atrial
fibrillation.
Study burden and risks
Risks associated with an ablation procedure. Risks associated with the
Cryoablation procedure do not differ from the standard ablation procedure.
Lambroekstraat (Green Square) 5D
Diegem 1831
BE
Lambroekstraat (Green Square) 5D
Diegem 1831
BE
Listed location countries
Age
Inclusion criteria
1. Subjects indicated for the treatment of AF with the cryoablation system
according to current and future Guidelines and system indications for use;
2. Subjects who are willing and capable of providing informed consent;
3. Subjects who are willing and capable of participating in all testing
associated with this clinical study at an approved clinical investigational
center;
4. Subjects whose age is 18 years or above, or who are of legal age to give
informed consent specific to state and national law.
Exclusion criteria
1. Any known contraindication to an AF ablation or anticoagulation, including
those listed in the instructions for use;
2. Subjects with indication for treatment of AF that is not according to
current and future Guidelines and system indications for use;
3. Atrial fibrillation secondary to electrolyte imbalance, thyroid disease, or
any other reversible or non-cardiac cause;
4. Known or pre-existing severe Pulmonary Vein Stenosis;
5. Evidence of myxoma, LA thrombus or intracardiac mural thrombus;
6. Previous cardiac surgery (e.g. ventriculotomy or atriotomy, CABG, PTCA,
stent procedure) within 90 days prior to enrollment;
7. Implantable cardiac device procedures (e.g. PM, ICD, CRT) within 30 days
prior to enrollment;
8. Implanted Left Atrial Appendage Closure device prior to the index procedure;
9. Interatrial baffle, closure device, patch, or patent foramen ovale (PFO)
occluder;
10. Subjects with severe valvular disease OR with a prosthetic - mechanical or
biological - heart valve (not including valve repair and annular rings);
11. Presence of any pulmonary vein stents;
12. Active systemic infection;
13. Vena cava embolic protection filter devices and/ or known femoral thrombus;
14. Any previous history of cryoglobulinemia;
15. History of blood clotting or bleeding disease;
16. Any prior history of documented cerebral infarct, TIA or systemic embolism
(excluding a post-operative deep vein thrombosis (DVT)) <= 180 days prior to
enrollment;
17. Subjects who are hemodynamically unstable;
18. The subject is unable or not willing to complete follow-up visits and
examination for the duration of the study;
19. Life expectancy <= 1 year per investigator*s opinion;
20. Women of childbearing potential who are, or plan to become, pregnant during
the time of the study (method of assessment upon investigator*s discretion);
21. Unrecovered/unresolved Adverse Events from any previous invasive Procedure;
22. Subjects who are currently enrolled in another investigational study or
registry that would directly interfere with the current study, except when the
subject is participating in a mandatory governmental registry, or a purely
observational registry with no associated treatments; each instance must be
brought to the attention of the sponsor to determine eligibility.
Design
Recruitment
Medical products/devices used
Followed up by the following (possibly more current) registration
No registrations found.
Other (possibly less up-to-date) registrations in this register
No registrations found.
In other registers
Register | ID |
---|---|
CCMO | NL73440.100.20 |