Monitoring a patient’s oxygen status during anesthesia using pulse oximetry is essential. The new Oxygen Reserve Index (ORI) may help provide clinicians real-time visibility to oxygenation in a moderate hypoxic range and serve as a warning of an…
ID
Bron
Aandoening
Oxygen Reserve Index
Ondersteuning
Onderzoeksproduct en/of interventie
Uitkomstmaten
Primaire uitkomstmaten
Validation of the new Oxygen Reserve Index (ORI) by collecting optical data using the Rainbow SET Pulse CO-Oximeter (Masimo Corp.) and comparing them to whole blood references of arterial and venous blood drawn in healthy volunteers
Doel van het onderzoek
Monitoring a patient’s oxygen status during anesthesia using pulse oximetry is essential. The new Oxygen Reserve Index (ORI) may help provide clinicians real-time visibility to oxygenation in a moderate hypoxic range and serve as a warning of an impending hypoxic state. In this way, ORI may enable proactive interventions to avoid hypoxia.
Onderzoeksopzet
during 4 hours
Onderzoeksproduct en/of interventie
•Change in inspiratory oxygen fraction (FiO2):
•Normoxia: room air (FiO2 0.21) (baseline)
•Hyperoxia: 21% to 100% oxygen (FiO2 0.21 to 1.0)
•Hypoxia: breathing mixtures of N2 in air via a tight fitting facemask to reach arterial oxygen saturation levels of 90%, 80%, and eventually 70% as measured by pulse oximetry (Radical 7, Masimo Corp.)
•Change in expiratory carbon dioxide concentration (etCO2) (during normoxia, awake state):
•Normoventilation (etCO2 4.5 – 5 Vol%)
•Partial rebreathing or adding external medical CO2 (target etCO2 7 – 7.5 Vol%)
•Hyperventilation (target etCO2 2.5 – 3 Vol%)
Algemeen / deelnemers
Wetenschappers
Belangrijkste voorwaarden om deel te mogen nemen (Inclusiecriteria)
• Age between 18 and 45 years
• Written informed consent
• Healthy subjects
Belangrijkste redenen om niet deel te kunnen nemen (Exclusiecriteria)
• Pregnant women (female subjects will have a pregnancy test prior to being admitted to the study).
• Presence of any cardiovascular or pulmonary disease